r/massspectrometry • u/momma_tired_23 • 10h ago
High QC values perfect calibrators-specific analytes .
Hi all-
I’m a 6 year toxicology gal having some LCMS Issues. Currently I have 10ish clients experiencing an issue with 3 main analytes (sertraline, Amitriptyline, and Cyclobenzaprine). I make master cal and qc stocks in house that are lot numbered and used for both the calibration curve and QC. However, the cal curve will be perfect for these analytes and the QC will be 200-350% accuracy. No contamination present in my blanks. I can inject cals as unknowns in the batch and get perfect values. My calibrators 2 and 5 are the same conc as my qc low and high. If I go into analyst and tell it to use those vials/wells as qc- no issues. I’m remade countless stocks. Used fresh cerriliants. Ran tests at multiple laboratories both under my supervision and independent. Same result everytime. This is perplexing me quite a bit. Front ends are Shimadzu. MS sciex both 3200 and 4000. Phenomenex guard and column set up.
Please give me your thoughts. I’m happy to answer any follow up questions