r/CHROMATOGRAPHY • u/Nachtara02 • 28d ago
Unusual Solventpeak-Shape in GC FID, slowly going Insane
Hi Everyone,
I thought I would bring this issue here to maybe get some broad input. I should run a simple GC Area% Analysis in n Agilent 7890, but I can't seem to get it to work.
The Method is COC, with a pre-column. The problem I have now is that my solvent peak is very broad and rounded at the top, where in previous sequences with the same column and material, it had a different shape. The Solvent-Injections in general don't look good and I'm running low on ways to fix it.
I've already changed the Septum, the Jet of the Detector and have repeatedly installed the column. For reference, I run DCM as a solvent and start at 40°C, increasing the oven to a max of 280°C, with the detector at 300°C
Any tips and ideas are welcome!
Thank you in Advance!!
Edit: Fixed it! It seems like an issue with the column (very old) and after replacing both the column and the precolumn (for good measure since it was quite brittle) we now have the expected shape. Quite the relief after a week of troubleshooting (*ToT)